FEI Helios NanoLab 600i DualBeam
Stanford Nano Shared Facilities
A DualBeam SEM/FIB combining a Ga+ focused ion beam and a high resolution field emission SEM column for milling, imaging, analysis and sample preparation at the nanoscale.
Hubrief prepared this page from public sources. It doesn't mean the organization has joined or endorsed Hubrief.
Capabilities
Focused ion beam
Milling with a focused Ga+ ion beam
As the source puts it: “milling, imaging, analysis, and sample preparation down to the nanoscale”
Scanning electron microscopy
High resolution field emission SEM imaging
As the source puts it: “high resolution field emission scanning electron ("Elstar") column”
Energy-dispersive X-ray spectroscopy
Elemental analysis with EDS detector
As the source puts it: “EDAX TEAM EDS Enhanced Analysis System including Octane Plus detector”
Automated TEM sample preparation
Automated TEM sample preparation with sectioning
As the source puts it: “AutoTEM Automated TEM sample preparation with sectioning”
Gas injection deposition
Tungsten, carbon and platinum deposition and etching gases
As the source puts it: “Gas Injection Systems”
Instrument & Specifications
Kind Electron microscopy
Equipment model FEI Helios NanoLab 600i
- Type
- DualBeam SEM/FIB system
- Manufacturer
- FEI
- Model
- Helios NanoLab 600i
- Configuration here
- Omniprobe AutoProbe 200 in situ lift-out, EDAX TEAM EDS with Octane Plus detector, retractable STEM detector, gas injection systems, FIB charge neutralizer, plasma cleaner, FEI CryoCleaner
As Stanford Nano Shared Facilities states for its unit
- Detectors
- Elstar in-lens SE detector (TLD-SE), in-lens BSE detector (TLD-BSE), Everhart-Thornley SE detector (ETD), retractable STEM detector with BF/DF/HAADF segments
- Spot size
- 4.0 nm at 30 kV using preferred statistical method
- Resolution
- 0.8 nm at 30 kV (STEM)
- Sample size
- Maximum sample size: 150 mm diameter with full rotation
- Accelerating voltage
- 0.35 – 30 kV
Access & Collaboration
Access not stated
- Booking
- Users must become qualified via nano@Stanford membership, Online SEM course, Helios Training Request Form; do not request training in NEMO
- Sample requirements
- Sample must withstand high vacuum without outgassing, be clean, electrically grounded; biological samples cannot exceed BioSafety Level 1
The official source doesn't say whether outside users can use it. Owning equipment doesn't mean it can be booked.